2015.14 Vitamin B (July 2018)
2015.14 (JULY 2018) BVit‐02 Total Bs ERP Evaluation Form
FOR ERP USE ONLY DO NOT DISTRIBUTE
Accuracy ‐ Analytical bias compared to reference material Certified Reference Material
Table 2
Measured Results
T stat Bias
Confidence interval
CRM product
Analyte
Certified value Uncertainty Certified range Concentration Difference from certified value
p ‐value
Nominal
990 987
90.1 92.8 38.8 76.0
1041 1004 1074 1067 1079 1117
6.3
13.3
SRM 1849a
Niacin (ug/100g)
1080
100
980 ‐ 1180
‐50.3
3.792
0.003
1000.5 ‐ 1058.9
1.4
‐37.3 100.9 103.1
979 977
1030 1012
49.8 68.5 50.3 46.0
Mean
1029.7
SD
46.0
d
T stat =
/ sd
n
where: = mean of the differences between measured results and reference value sd = standard deviation of the differences between results and reference value n = number of replicate analyses of SRM d
H 0 : = 0, The null hypothesis is that there is no difference between the measured results and the certified value:
d
p ‐value: is the probability (at the 95% levels of confidence) of obtaining the test statistic assuming that the null hypothesis is true p ‐value < 0.05 indicate a difference of statistical signifance between measured results and SRM.
95% confidence interval: The true population of the measured value expected to fall within the confidence interval estimate 95% of the time
sd
sd
t crit
t crit
lower limit = mean ‐
upper limit = mean +
n
n
crit t
where: = critical values of the t‐distribution
sd = standard deviation of the differences between results an reference value n = number of replicate analyses of SRM