2018 August 26 SPSFAM Book
Table 1. Method Performance Requirements
Residue Concentration in Matrix
N Acceptance Criterion
0 (blank)
30 ≤10% POD with 95% confidence
≤ (or ≥ 90%) 10% POD with 95% confidence ≥90% POD with 95% confidence b
0.5x MRL
30 per drug a 30 per drug a
1.0x MRL
a Tested as drug cocktail(s). b All incorrect results must be investigated for the determination of the concentration at which POD 90 with 95% confidence is achieved.
System suitability tests and/or analytical quality control: • Suitable methods will include blank check samples and check standards at 0.5 x MRL and 1.0 x MRL.
• Method developers will provide information on how cutoffs were determined. • Method developers may prepare cocktails of multiple drug residues.