Reddy: J ournal of AOAC I nternational V ol. 98, N o . 4, 2015 1137
( 2 ) Transfer to suitable storage vessel (minimum 25 mL capacity with a wide mouth). ( 3 ) Add 24 mL 3% pyridine in ethanol and mix well. Expiration 5 days. ( n ) Mobile phase buffer .—200 mM ammonium acetate in water. ( 1 ) Weigh 1.54 ± 0.05 g ammonium acetate in weighing pan. ( 2 ) Transfer to 100 mL volumetric flask with water. ( 3 ) Add 50 mL water and swirl until dissolved. ( 4 ) Fill to volume with water and mix well. Expiration 5 days. ( o ) Mobile phase A .—10 mM ammonium acetate in water. ( 1 ) Transfer 25 mL mobile phase buffer to 500 mL mobile phase bottle. ( 2 ) Add 475 mL water and mix well. ( 3 ) Sonicate for 5 min. Expiration 5 days. ( p ) Mobile phase B .—10 mM ammonium acetate in acetonitrile–water (95 + 5, v/v). ( 1 ) Transfer 25 mL mobile phase buffer to 500 mL mobile phase bottle.
Table 2015.04A. Preparation of calibration standards
Working standard (1 + 10), μL
Working internal standard, μL
Working standard, μL
Acetonitrile–water (60 + 40, v/v), μL
Standard
1 2 3 4 5 6
10 20 50
40 40 40 40 40 40
950 940 910 950 910 860
10 50
100
( 3 ) Add 50 mL ethanol and mix well. ( k ) Pyridine, 3% in ethanol. ( 1 ) Transfer 3 mL pyridine to 100 mL glass bottle. ( 2 ) Add 97 mL ethanol and mix well. ( l ) 2-NPH .—Approximately 0.1 M in 0.1 N HCl–ethanol (50 + 50, v/v). ( 1 ) Weigh 0.375 ± 0.010 g wet powder into 25 mLvolumetric flask. ( 2 ) Add ~20 mL 0.1 N HCl–ethanol (50 + 50, v/v). ( 3 ) Sonicate and swirl until dissolved. ( 4 ) Fill to volume with 0.1 N HCl–ethanol (50 + 50, v/v). Mix well. Expiration 5 days.
( 2 ) Add 475 mL acetonitrile and mix well. ( 3 ) Sonicate for 5 min. Expiration 5 days. ( q ) Calibration standards (STD 1–6) .
( 1 ) Add volume aliquots (µL) listed in Table 2015.04A to glass centrifuge tubes. For method blank, use 1 mL acetonitrile– water (60 + 40, v/v). ( 2 ) Vortex.
( m ) EDC. —0.25 M in 3% pyridine in ethanol. ( 1 ) Weigh 1.20 ± 0.05 g EDC in weighing pan.
Table 2015.04B. Instrumental conditions
Chromatography
Flow rate
300 µL/min
Column temperature
40 ° C 10 μL
Injection volume
Sample temperature Gradient program
Ambient
Time 0.00 1.00 7.00 9.00 9.01 12.00
% A
% B
Curve
90 90 45
10 10 55
6 6 6 6 6 6
0
100
90 90
10 10
MS tune
Ionization mode
ESI-
Cone gas Nebulizer
250 L/h 7.0 bar
Capillary
0.5 kV 20.0 V 150 ° C 350 ° C 900 L/h
Source offset
Collision gas flow Quad 1 resolution Quad 2 resolution MS calibration range
0.15 mL/min
Source temperature Gas temperature Desolvation gas flow
Unit mass (0.75 Da FWHM) Unit mass (0.75 Da FWHM)
50–2000 amu
MS/MS transitions
Compound MFA-2NPH
Parent mass, m/z Daughter mass, m/z
Dwell, s
Cone (V)
Collision energy , V
212 216
182 186
0.15 0.15
20 20
15 15
MFA-2NPH_IS