Method Performance Requirements
Required Analytes: • Arsenic • Cadmium • Mercury • Lead
Table 1: (optional) • Antimony • Barium • Chromium • Copper
• Nickel • Silver • Zinc
9
Method Performance Requirements
Limit of Quantitation (LOQ)
≤ 10 ppb ( μ g/kg)
Reproducibility (RSD R )
Repeatability (RSD r )
Range
Recovery
>100 ppb to 1 ppm ≥ 10 ppb to 100 ppb > 1 ppm to 10 ppm
15% 11% 7.3%
32% 16%
60% - 115% 80% - 115% 80% - 115%
8%