4. Method Performance Requirements :
Limit of Quantitation (LOQ)
≤ 10 ppb (μg/kg)
Reproducibility (RSD R )
Repeatability (RSD r )
Range
Recovery
>100 ppb to 1 ppm ≥ 10 ppb to 100 ppb > 1 ppm to 10 ppm
15% 11% 7.3%
32% 16%
60% - 115% 80% - 115% 80% - 115%
8%
58 59 5. System suitability tests and/or analytical quality control: 60
61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91
Suitable methods will include blank check samples, and check standards at the
lowest point and midrange point of the analytical range.
6. Reference Material(s):
A certified reference material should be used when available. Internally produced reference materials may be used for a variety of cannabis products such as plant material, concentrates, and finished products until reference material is made available by an internationally-recognized organization such as Institute for Reference Materials and Measurements (IRMM) or United States National Institute of
Standards and Technology (NIST).
7. Validation Guidance & References :
Recommended level of validation: Official Methods of Analysis SM
USP <233>
8. Maximum Time-To-Result : No maximum time.