Comments Submitted
Comment or Concern
Response
Edits: (1) Capitalization of residual solvents is inconsistent in Table 1 (e.g., "Acetonitrile" and "acetonitrile" and "Acetone" and "acetone") can be found in Table 1. (2) Duplicates were removed; lowest action levels are listed.
Table 1: Why are some residual solvents listed more than once and with different limits? Some explanation should be offered.
Line 34 : What is signal and what is noise? Apparent concentration is signal; noise is noise of blank Recommendation to put “background” before noise. Line 55: Recommendation to specify short period on….. … “individually prepared test portions”
Edit: SMPR to "signal to background noise"
Edit: Variation arising when all efforts are made to keep conditions constant by using the same instrument and operator and repeated on individually prepared sample/test portions.
Comments Submitted Comment or Concern Table 1: Add less than or equal to at all lower levels in each row. Recommendation to reduce lower end of analytical range to ½ of the target LOQ (or 1/3) Table 3: 1 – How do we verify these are met? Suggestion to reference appropriate AOAC document. Note: review the RSDR in light of comment regarding feasibility as discussed by Robert LaBudde. Suggestion to compare observed RSDR to limit. Make units in this table consistent (will use PPM)
Response
Edits: (1) Changed: Lower (≤ ppm) (2) Changed target LOQ to 50% of Action Level (1) These are referenced in AOAC Appendix F
(2) SMPR follows AOAC Appendix F guidelines
(3) SMPR follows AOAC Appendix F guidelines Edit: (4) mg/kg was replaced with ppm