AOAC Final Action Methods in 2017
J OSEPH ET AL . : J OURNAL OF AOAC I NTERNATIONAL V OL . 101, N O . 3, 2018 841
Table 8. Comparison with LC-MS/MS technique
Biotin, µg/100 g
AOAC Method 2016.02
Sample No.
Product description
Blind duplicate
LC-MS/MS Difference
MLT 1
Infant formula powder, partially hydrolyzed milk-based
A B A B A B A B A B A B A B A B A B A B A B A B
33.78 34.23 78.79 79.61
35.55 39.10 82.45 81.15
– 1.77 – 4.87 – 3.66 – 1.54
MLT 2
Infant elemental powder
Infant formula RTF, milk-based a
MLT 3
4.02 4.13
3.73 3.84
0.29 0.29 5.84
MLT 4
Adult nutritional RTF, high-fat
71.90 71.02 26.67 27.27 44.36 45.14
66.06 77.49 26.99 25.25 43.00 42.27
Downloaded from https://academic.oup.com/jaoac/article-abstract/101/3/831/5653969 by guest on 05 May 2020
– 6.47 – 0.32
MLT 5
Infant formula powder, milk-based
2.02 1.36 2.87 9.48
MLT 6
Infant formula powder, soy-based
MLT 7
NIST SRM 1849a
196.06 196.96 260.88 256.49 174.60 159.09
186.58 181.50 263.90 246.17 171.69 162.77
15.46 – 3.02 10.32
MLT 8
Adult nutritional powder, low-fat
MLT 9
Child formula powder
2.91
– 3.68
MLT 10
Toddler formula powder, milk-based
10.38
9.87
0.51
9.85
10.30 41.66 39.77 54.05 51.57
– 0.45
MLT 11
Infant formula powder, milk-based
42.84 42.96 52.63 52.73
1.18 3.19
MLT 12
Adult nutritional RTF, high-protein
– 1.42
1.16
a
RTF = Ready-to-feed.
technique was not given. Rather than performing a true backflush, the 3 mL methanol eluate was recycled through the IA cartridge three times. It is possible that this may partly explain the slightly low recovery for some samples, given that the IA cartridge may well have retained a small portion of this eluent fraction/biotin. The details of the backflush technique using a syringe plunger and reverse pressure are now detailed in the method.
IAC Loading Speed
Two laboratories obtained lower biotin results for the practice samples initially and were advised to follow the IAC loading by gravity, as documented in the method. The results came back higher and comparable to other laboratories.
Elution of Biotin from IAC
Syringe Filter
While assisting a participant investigating low biotin results, it became apparent that the backflush technique was not exactly followed, mainly, it seems, because the exact description of this
At least two laboratories experienced interference with chromatography. The investigation isolated the issue as the
Table 9. Comparison of method performance with SMPR 2014.005
AOAC 2016.02
Parameter
SMPR 2014.005
Comments
Analytical range, µg/100 g
0.1 – 150
0.1 – 300
Analytical range of the proposed method is wider based on the sample weight and loading volume for IAC extraction The proposed method meets the LOQ of 0.1 µg/100 g as required by SMPR 2014.005 Repeatability of the proposed method is better than the SMPR 2014.005 Reproducibility of the proposed method is better than the SMPR 2014.005
LOQ, µg/100 g
≤ 0.1
≤ 0.1
≤ 6% at >1 µg/100 g
≤ 5% at >1 µg/100 g
RSD r
≤ 12% at >1 µg/100 g
≤ 10% at >1 µg/100 g
RSD R
HorRat
0.5 – 2.0 90 – 110
0.3 – 0.5 95 – 105
Excellent precision performance rating
Recovery of >1 µg/100 g, %
The spiked recoveries by the proposed method are better than the recovery range specified by SMPR 2014.005
NIST SRM 1849a, µg/100 g
199 ± 13
197 ± 10
NIST 1949a results are within the certified limits