Table 2015.10C. LC-MS parameters
LC-MS settings a
LC-MS model
Sciex 4000
Sciex 6500
Ionization mode IonSpray voltage
MRM
MRM
1000 V 550 ° C
2000 V 550 ° C
Turbo IonSpray temp. Entrance potential (EP) Collision gas (CAD) Curtain gas (CUR) Nebulizing gas (GS1) Nebulizing gas (GS2)
10 V
10 V
Nitrogen, 5.0 Nitrogen, 20.0 Nitrogen, 60.0 Nitrogen, 60.0
Nitrogen, 5.0 Nitrogen, 20.0 Nitrogen, 60.0 Nitrogen, 60.0
Needle position
Y = 5 mm, X = 5 mm
Y = 5 mm, X = 5 mm
LC-MS model
Agilent 6490
Ionization mode
MRM 380 V
Fragmentor
Cell accelerator
5 V
Gas temp. Gas flow Nebulizer
200 ° C
20 L/min
30 psi 225 ° C
Sheath gas temp. Sheath gas flow
11 L/min 1500 V
Capillary +
Nozzel voltage + High-pressure RF + Low-pressure RF + Ionization mode Capillary voltage RF lens voltage Source temp. Nebulizer temp. LC-MS model
500 V 200 V 100 V
Waters TQD
MRM
2000 V 0.1 eV 150 ° C 450 ° C
API gas
On
Nebulizer gas Desolvation gas
Nitrogen
900 L/h nitrogen 50 L/h nitrogen 0.25 mL/min argon
Cone gas
Collision gas
API gas
On
Nebulizer gas
Nitrogen
Transition monitored Q1/Q3 b
Analyte
Transition use/type
Dwell time, msec b
Collision energy, V b
Retention time, min
Carnitine
Quantitation
162.0/103.0 165.0/103.0 162.0/84.4 162.0/59.1 104.2/60.0 104.2/42.2 108.2/60.0
100 100 100 100 100 100 100
25 25 29 27 25 25 25
1.7 1.7 1.7 1.7 2.1 2.1
Carnitine-d3
Internal standard
Carnitine Carnitine Choline Choline
Qualifier Qualifier
Quantitation
Qualifier
Choline-d4 2.1 a MS/MS settings may be modified except for ionization type and mode to obtain optimum chromatography and sensitivity. Exact mass ions may vary slightly from instrument to instrument because of unit resolution of quadrupole mass spectrometers. b Settings listed are for Sciex series of LC-MS. Precursor/product masses and instrument settings may vary. Internal standard