AOAC Guidance on FA Immunoassay Validation (August 2023)
538 539 540 541 542 543 544 545 546 547 548 549 550 551 552 553 554 555 556 557 558 559 560 561 562 563 564 565 566 567 568 569 570 571 572 573 574 575 576 577
intermediate precision. Data used must meet other method performance
criteria (e.g., recovery of 50-150%).
5.2.4.2.5. Recovery Assessment
5.2.4.2.5.1. Data collected for the purposes of precision evaluation may also be used for
the recovery assessment.
5.2.4.2.5.2. If conducted separately from the precision assessment, evaluate each incurred matrix with six independent analyses (test portions) per concentration level at a minimum of three non-blank concentration levels
covering the analytical range.
5.2.4.3. Data Analysis and Reporting
5.2.4.3.1. Precision
5.2.4.3.1.1. Nested Designs: Repeatability and Intermediate Precision
5.2.4.3.1.1.1. Data generated from nested designs, such as those as described above, should be analyzed by an ANOVA capable of providing estimates of
intermediate precision and repeatability.
5.2.4.3.1.1.2. Annex C contains full instructions, R code, and example datasets for the
study designs described in this guidance.
5.2.4.3.1.2. Repeatability Only
5.2.4.3.1.2.1. In a situation where a study design for estimating repeatability alone is selected, the mean, standard deviation, and relative standard deviation should be calculated for each test material (i.e., each matrix- concentration combination). Formulas for standard deviation and relative standard deviation, as defined in Appendix F (3), are as follows:
Repeatability standard Deviation (s r ): s r = [ Σ (x r – x̅ )
2 /n] 0.5
Repeatability relative standard deviation (RSD r ): RSD = s r × 100/ x̅
5.2.4.3.1.2.2. The study report must include the standard deviation and RSD values for each test material, and all repeatability estimates must meet requirements set forth in the relevant SMPR or established by the ERP. In the absence of an SMPR and ERP, acceptable RSD r values for food allergen
immunoassays are generally ≤20%.
5.2.4.3.2. Sensitivity: LOD, LOQ
5.2.4.3.2.1. LOD will be estimated using a hypothesis test approach, with α = β = 0.05. The relationship between observed concentration and intermediate precision standard deviation must be taken into account in the estimation of LOD (also referred to as a precision profile estimation method for LOD). Full instructions
for the calculations to estimate LOD are in Annex C.
5.2.4.3.2.2. LOQ estimation will be based on the relationship between concentration and
intermediate precision standard deviation. Full instructions for the
calculations to estimate LOQ are in Annex C.
5.2.4.3.3. Recovery
Made with FlippingBook - Share PDF online