AOAC ISPAM Stakeholder Panel Meeting Book 3-13-18
103 104 105 106 107 108 109 110 111 112 113 114 115
Appendix D: Guidelines for Collaborative Study Procedures to Validate Characteristics of a Method of Analysis; 20 th Edition of the Official Methods of Analysis SM of AOAC INTERNATIONAL (2016). Available at: http://www.eoma.aoac.org/app_d.pdf Appendix F: Guidelines for Standard Method Performance Requirements; 20th Edition of the Official Methods of Analysis SM of AOAC INTERNATIONAL (2016). Available at:
http://www.eoma.aoac.org/app_f.pdf
Appendix M: Validation Procedures for Quantitative Food Allergen ELISA Methods: Community Guidance and Best Practices; 20th Edition of the Official Methods of Analysis SM of AOAC INTERNATIONAL (2016). Available at: http://www.eoma.aoac.org/app_m.pdf
Table 1: Method performance requirements
Parameter
Minimum Acceptance Criteria for target matrix
Lower limit Upper limit
≤10
Analytical Range (ppm) 1
≥ 20 2
LOQ (ppm) 2
≤10
LOD (ppm) 2
≤10
50-150%
Recovery (%) 3
% RSD
≤20%
r
% RSD
≤30%
R
Notes: 1 ppm in peanut.
2 See “Choice of LOD / LOQ for Quantitation of Milk by ELISA-based Methods” (http:\wwww.aoac.org\xxyyzz) memo for rationale for setting lower limit of range. 3 Use incurred samples as per Appendix M. Incurred materials can be obtained from MoniQA Association.
116 117
1 . Ibid. 2 . Ibid. 3 . ISO 5725-1:1994 4 I b id.