AOAC RI ERP EBOOK FOR FERTILIZERS
for instrument drift. The inclusion of a digestion blank, a sample duplicate, and known reference materials is highly encouraged. I. Calculations
volumetric flask, dilute to volume with H 2 acid-washed polypropylene bottle. ( h ) Sampler wash solution, 1% HNO 3
O, mix, and transfer to an
.—Dilute 10 mL of trace
metal grade HNO 3
to 1000 mL with H 2
O.
Element, mg / kg = (instrument concentration, mg / L)(100 mL)(1 L / 1000 mL) (sample weight, g)(1 kg / 1000 g)
H. Determination Analyze test solutions using an ICP-OES instrument calibrated with standard solutions. Insert a 10 mg/kg working standard or other suitable quality control solution every 10 test portions, to monitor
where 100mL assumes themicrowave digest is diluted to 100mL. Reference: J. AOAC Int. 89, 1447(2006) .
AOAC Research Institute Expert Review Panel Use Only
ã 2009 AOAC INTERNATIONAL
Made with FlippingBook