8. ADDITIONAL REAGENTS
109
110
Lysis reagent – Using test kit reagents, pipette 150 µ L of protease into one 12-mL bottle of lysis
111 112 113
buffer. Label bottle with the date prepared. Reagent will remain stable for up to two weeks if
stored at 2-8ºC.
9. APPARATUS
114
115
Incubators – For maintaining media at 35 ± 1ºC and 39-42ºC.
116 117
Stomacher – Seward model 400 or equivalent for mixing the sponge sample with enrichment
media.
118
BAX ® System apparatus
119
BAX® System Q7 cycler/detector with computer workstation
120
BAX® System application software
121
Cluster tubes with caps and racks for lysis
122 123 124 125
Capping/de-capping tools – for removing and sealing cluster tube caps and PCR
tube caps without jarring the contents
Heating and cooling blocks with inserts – for maintaining lysis tubes at 37±2ºC,
95±3ºC, and 2-8°C.
126
AOAC INTERNATIONAL Pipettes – for transferring reagents; two adjustable mechanical pipettes covering 20- 200 µ L and 5-50 µ L; one repeating pipette; and one multi-channel pipette covering 8 channels and 5-50 µ L. Pipettes should be calibrated to deliver required volumes
127
128 129 130 131
within 10%.
Pipette tips with barriers: 0.5-250 µ L, 0.5-100 µ L extended barrier; 5 mL repeater
pipette tips