Dupont BAX Salmonella PTM Report Modification Approved 2012 / PTM Certification No. 081201 For Expert Review Panel Use Only Do Not Distribute
307 308
14. METHOD COMPARISON RESULTS
309
310 311 312 313 314
The results of the method comparison studies are summarized in Tables 2 and 3 below. For
all sample types and enrichment methods tested, the updated BAX® System method
returned results that were identical to the certified method. POD analysis demonstrates no
statistically significant difference between results of the updated BAX® System method and
the reference culture methods.
315
Fluorescent Probe Performance
316
317 318 319 320 321 322 323
Performance of the fluorescent probes in the additional eight cycles do not appear to suffer
either by photo bleaching, which would lead to decreased signal for a positive reaction, or by
quencher degradation, which would lead to a false positive rise in signal. Thousands of reactions
have been run internally to date in the algorithm training and validation phases of the modified
protocol. Examples of typical reactions are shown below in Figure 1 (negative reactions) and
Figure 2 (positive reactions). Data in these figures is typical of data obtained during algorithm
training. Additional data is available upon request.