4. Conclusion.

The method used for this SLV meets the requirements as described in the AOAC SMPR 2014.002 for fructans in Infant Formula and Adult/pediatric Nutritional Formula in which fructan is present in a concentration ranging from 0.03 to 0.5 g/100g reconstituted final product. The precision as done during the SLV meets the repeatability as proposed by the workgroup, being ≤6%. The accuracy was performed by spiking the Test Materials kit samples by 3 different pure fructan products of different origin (sugar and chicory) and DP range. The spiking/recovery is ranging from 96.9 to 107.1% with a mean of 100.3%, which is very well meeting the proposed requirement of 90-110%. The very low concentrations (spike level 0.05) are included during this spiking test and even these meet the proposed range.

Annex 1 Chromatogram of a standard. Annex 2 Calibration curve for fructose and sucrose. Annex 3 Chromatogram of a sample J4G0975 (Infant elemental powder 00403RF00). Annex 4 Description of the GC method.

Tiense suikerraffindarij/ Beneo-Orafti (part of Südzucker group) Central Department Research, Development and Services-Analytics-Tienen Monique Steegmans