HMO-02 (Analysis Paper) 10-2022 FOR ERP USE ONLY DO NOT DISTRIBUTE
Calibration fit A linear model forced through zero was used to fit the calibration data for all HMOs (Figure 5 - Figure 11), and the model appears to be a good fit. The relative residuals were calculated and plotted against the analyte concentration (Figure 5 - Figure 11), for all but two points the predicted concentration is within 5 % of the actual concentration. The two points beyond that limit were both for 6’SL, one point at 3.73 µg/ml (-7.3 %), and one point at 74.7 µg/mL (+5.9 %). For 2’FL and 3FL the plots of the relative residuals (Figure 5 and Figure 6) are not randomly distributed, there is a tendency for underestimation at low concentrations. However, the magnitude of the deviation is small enough that it does not have a significant impact on the analytical results.
B
A
Figure 5: Calibration fit for 2'FL. A: Calibration curve. B: Plot of relative residuals
A
B
6.0
4.000
3.500
y = 0.0055x R² = 0.9998
4.0
3.000
2.0
2.500
2.000
0.0
0
100 200 300 400 500 600 700
1.500
-2.0
1.000
Concentration (%)
Peak Area Ratio (3FL/IS)
0.500
-4.0
0.000
Difference Berween Actual and Calculated
0
100 200 300 400 500 600 700
-6.0
3FL Concentration (µg/mL)
3FL Concentration (µg/mL)
Figure 6: Calibration fit for 3FL. A: Calibration curve. B: Plot of relative residuals
20