AOAC SPIFAN Nutrients ERP (December 7, 2022)

Amino-06 (SLV) 10-2022 FOR ERP USE ONLY DO NOT DISTRIBUTE

sample at 3000 rpm for 30 minutes. Filter the supernatant through a 0,2 µm syringe filters into an injection vial. d. System blank – Pipet 400 µL UHQ water in a 15 mL centrifuge tube and dilute by addition of 1600 µL acetonitrile. Homogenize the system blank and filter through a 0,2 µm syringe filter into an injection vial. e. Method blank – Weigh 2,5 grams of UHQ water in a 50 mL volumetric flask. Pipette 100 µL of IS solution (E(b)) and 2 mL of 50% TCA solution (E(c)) to the volumetric flask. Fill to the mark and homogenize. Let the samples deproteinize for 10 minutes. Filter the deproteinized samples over filter paper, discard the first 5 mL Pipet 400 µL filtrate in a 15 mL centrifuge tube and dilute by addition of 1600 µL acetonitrile. Homogenize the diluted sample and subsequently centrifuge the sample at 3000 rpm for 30 minutes. Filter the supernatant through a 0,2 µm syringe filter into an injection vial. f. Calibration Standard blank (T0) – Pipet 100 µL of IS stock solution (E(b)) into a 50 mL volumetric flask. Fill to the mark with UHQ water and homogenize. Pipet 400 µL from the flask in a 15 mL centrifuge tube and dilute by addition of 1600 µL acetonitrile. Homogenize the standard blank and filtrate through a 0,2 µm syringe filter into an injection vial.

I.

Chromatographic Conditions

The separation of taurine is performed by a UHPLC system equipped with a HILIC column. The column temperature is kept at 45±1° Celsius. The injection volume is 3,5 µL and the autosampler temperature is 15±1°C. Taurine is separated according to the gradient presented in table 3. Detection of the component of interest is performed by a mass spectrometer.

Table 3. Gradient program

Eluent

Time (min)

Ammonium formate (A (%))

Acetonitrile (B (%))

Flow (mL/min)

0,0 4,5 4,5 5,5 5,5 8,0

4

96 51 14 14 96 96

0,4 0,4 0,4 0,4 0,4 0,4

49 86 86

4 4

J.

MS conditions

The detection of taurine and the IS is performed on a triple quadrupole MS in MRM mode. The settings describes for the setup of AB SCIEX MS are listed in table 4. Ionization is performed in a turbo spray ion source in positive mode and the data acquisition rate is 0,465 s with 1032 collection cycles. In table 4 the ion source parameters are described. The MRM transition parameters of taurine and the IS are defined in table 5. The MS settings in other laboraties should be optimized locally.

Made with FlippingBook - Share PDF online