AOAC SPIFAN Nutrients ERP (December 7, 2022)
Amino-07 (Method) 10-2022 FOR ERP USE ONLY DO NOT DISTRIBUTE
Table XXXX.XXC. Mass spectrometer instrument settings a Parameter
Value
ESI +
Ionization mode
Curtain gas
30 40 40
Nebulizer gas GS1
Heater gas GS2 Collision gas
N 2 (medium)
Source temperature Ion spray voltage
300°C
5500 V a These settings are suitable for the 6500 triple quadrupole mass spectrometer (Sciex); optimal settings on alternative instruments may differ.
(b) The specific compound parameters to be used are:
Table XXXX.XXD. Compound parameters
Precursor ion,
Product ion,
DP
b ,
EP c ,
CE d , CXP e , Dwell time,
Ion a
V
V
V
V
ms
m/z
m/z
quantifier qualifier
126.0 126.0 129.0
108.0
15 49 15
16 16 16
150 150 150
65.1
1
10
internal standard
111.0
a quantifier = taurine quantifier ion, qualifier = taurine qualifier ion, internal standard = 13 C 2
15 N taurine
ion. b DP = declustering potential. c EP = entrance potential. d CE = collision energy. e CXP = collision cell exit potential.
I.
Calculations (a) Concentration of 13 C 2 15 N taurine in stable isotope labelled stock standard ( SILSS ): SILTSS = 1 1 0 0 0 × 1000 where: SILSS Tconc = concentration of 13 C 2 15 N taurine in SILSS (μg mL −1 ) 10 = mass of 13 C 2 15 N taurine in SILSS (mg)