121 122 123
Table 3: Method performance requirements
Minimum acceptance criteria
Parameter
Analytical range
10 ppb
1 ppm*
–
Limit of quantitation (LOQ)
< 10 ppb*
10 – 100 ppb
60 – 115 %
Recovery
≥ 100 ppb – 10 ppm
80 – 110 %
≤ 10 – 50 ppb
≤20 %
Repeatability (RSDr)
≥ 50 – 300 ppb
≤13 %
≥ 300ppb – 1 ppm
≤12 %
≤ 10 – 50 ppb
≤30 %
Reproducibility (RSDR)
≥ 50 – 300 ppb
≤20 %
≥ 300 ppb – 1 ppm
≤18 %
* Arsenic (As) species and/or total inorganic arsenic in selected food and beverage matrixes.
124 125 126 127 128 129 130 131 132 133 134 135 136 137 138 139 140 141 142
4
Made with FlippingBook