AOAC2016-03ERPRECOMMENDATION
2016.03 (Nov. 2017) - Chlor-02/04 Final Action Review FOR ERP USE ONLY DO NOT DISTRIBUTE
= the volume of AgNO 3 solution consumed at titration endpoint for blank (G. d) (in ml); = the dilution factor for preparation of reconstituted powder, RTF or concentrate, For samples requiring the protein precipitation step (F. d-f) an additional factor (e.g. for 25 g sample f = 2) is needed;
V 0
f
35.45 = chloride mass in µg corresponding to 1 ml of 1 mol/l AgNO 3 ; 100 = the mass conversion to mg/100 g.
J.PRECISION
(a). General Details of the interlaboratory test of the precision of the method are summarized in Table 1. . The values derived from the interlaboratory test may not be applicable to analyte concentration ranges and/or matrices other than those given in Table 1.