BPA SMPR V 7
49 50 51 52 53 54 55 56 57 58 59 60 61 62 63
Repeatability
Variation arising when all efforts are made to keep conditions constant by using the same instrument and operator, and repeating during a short time period. Expressed as the
repeatability standard deviation (SD r
); or % repeatability relative standard deviation
(%RSD r
).
Reproducibility
The standard deviation or relative standard deviation calculated from among-laboratory data. Expressed as the reproducibility relative standard deviation (SD R ); or % reproducibility
relative standard deviation (% RSD R ).
4. Method Performance Requirements:
Table 1
LOD
≤ 0.1 µg / liter
LOQ
≤ 0.5 µg / liter
Table 2 Analytical range*
< 2 µg / liter
2-5 µg / liter 80% - 120%
5-20 µg / liter 80% - 120%
Accuracy
60% - 140%
%RSD r
≤ 20%
≤ 10% ≤ 20%
≤ 5%
%RSD R
≤ 40%
≤ 10%
Units are expressed as µg / liter as weight / volume. *Concentration in the ready to drink product
64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83
5. System suitability tests and/or analytical quality control:
Suitable methods will include blank check samples, and check standards at the lowest point
and midrange point of the analytical range.
6. Reference Material(s):
ANNEX F: Development and Use of In-House Reference Materials in Appendix F: Guidelines for Standard Method Performance Requirements in the Official Method of Analysis of the
AOAC INTERNATIONAL compendium.
7. Validation Guidance :
Method developers must submit data (LOQ, Accuracy…) on at least one of the matrices in Table 3. Data from as many of the other beverages listed in Table 3 would be desirable. It is
desirable to also have data on products containing coffee/dairy.
RSDRs may be calculated from pooled results from the different matrices in Table 3. Developers should submit the method’s procedures used for background assessment and
control, and frequency of analysis of method blanks.
2
Made with FlippingBook flipbook maker