NOTE2 As of 2017, many things are still not known about the effects of prolonged exposure to strong magnetic 497 field on the human body; however, epidemiological studies and animal experiments have raised the possibility 498 that damage can occur. With regard to demonstrated effects on humans, there are known cases of pacemaker 499 malfunctions in magnetic fields of 0,5 mT; therefore, the boundaries of the zone around the superconducting 500 magnet where the magnetic field is ≥0,5 mT shall be clearly indicated. 501 502

503 504 505

Figure 1 ‐

Example of a standard 400 MHz NMR apparatus installation

fan (with shutter and switch)

ceiling

ceiling

ceiling

duct

quench duct

duct

burst valve

outdoor

outdoor

outdoor

Figure 2 - Example of a standard NMR superconducting magnet and the ventilation/exhaust duct layout 506 Fig.2 Example of a standard NMR apparatus superconducting magnet and the ventilation/ exhaust duct layout