METHOD FOR THE ESTIMATION OF TOTAL PHENOLIC CONTENT USING THE FOLIN- C ASSAY
Results of validation: Following test were conducted to verify the performance of the method as per the requirement of AOAC SMPR 2015.009.
10.0
10.1. Analytical Range: Analytical range was demonstrated on reference standard. Different concentrations ranging from 5-100ppm were prepared and measure the absorbance.
Sr.no.
Concentration (ppm)
Absorbance
1
5
0.101
2
25
0.325
3
50
0.651
4
100
1.273
Calibration Curve of Gallic acid
1.4
1.2
1
y = 0.012x + 0.028 R² = 0.999
0.8
0.6
Abosrabance
0.4 Absorbance
0.2
0
0
50
100
150
Concentartion in ppm
Conclusion: The Co-relation coefficient (R 2 ) of 0.999 confirms that the concentrations ranges are linear and follows the Beers lambert’s laws.
10.2. Limit of Quantification (LOQ): Based on Linearity study a concentration of 5 ppm is found to be the LOQ. Six different preparation of 5ppm were analyzed and the relative standard deviation (RSD) is within 10%. Confirming LOQ to be 5ppm 10.3. Recovery: Recovery study was conducting by adding known amount (80 to 120% of working range) of standard in both the raw material & extract. The samples were analyzed using the methodology mentioned above (7.4). The recovery is within 80 to 110%.