Microbiology Methods for ERP Review 3-2020

65

GENE ‐ UP ® SLM 2 Collaborative Study

July 2019

c) Clip the GENE - UP® Lysis Tube Holder on the GENE - UP® Heavy Rack Holder. d) Transfer 20 μ L of sample into the lysis tube. Use the Plate Map from the GENE - UP® Routine software to pipet each sample into the correct plate position. e) Remove the GENE - UP® Lysis Tube Holder from the GENE - UP® Heavy Rack Holder. f) Clip the GENE - UP® Lysis Tube Holder on the Troemner Vortex Mixer Adaptor. g) Run the vortex mixer at 2200 rpm for 5 min. Note: When using the Vortex - Genie® Pulse, fit it with the GENE - UP® Lysis Rack Adaptor. Run the vortex at maximum speed for 5 min. The maximum speed must be above 2000 rpm. h) When lysis is complete, remove the GENE - UP® Lysis Tube Holder from the Troemner Vortex Mixer Adaptor. i) Clip the GENE - UP® Lysis Tube Holder into the GENE - UP® Heavy Rack Holder and proceed to Final Setup for PCR. Note: The lysate can be stored for up to 3 days at 2–8°C or at  ‐ 15 to  ‐ 31°C for extended storage. j) PCR Preparation a) Before beginning the procedure, don a clean pair of powder ‐ free latex or nitrile gloves. b) Use the plate map created in the GENE - UP® Routine software to determine the number of PCR tubes required from the GENE - UP® PCR Kit and place the correct number of PCR tubes in the GENE - UP® PCR Tube Holder. If less than eight tubes in a strip are required, the strips can be cut apart, and only the used tubes are placed in the GENE - UP® PCR Tube Holder. Note: Only remove the required number of strips from the pouch and carefully reseal the pouch after opening. c) Use the following steps to remove the transportation caps from the strips: d) Tap on the strips on the bench to ensure the pellets are on the bottom of the tubes. e) Carefully open the caps to prevent spilling the freeze - dried pellet. f) Visually check that the freeze - dried pellets are present at the bottom of each tube. g) Using a 10 μ L Biotix filter pipette tip with a single or multichannel pipette, transfer 10 μ L of lysed sample (red) in the appropriate PCR tube. To determine the appropriate plate position for each sample, refer to the Plate Map from the GENE - UP® Routine software. Note: Do NOT agitate the lysate before aspirating the sample. The solid material must stay at the bottom of the tube. Note: Visually check the tips to confirm the absence of beads, bubbles, and for correct volume of lysate. Note: For negative control procedure, use 10 μ L of control buffer instead of lysed sample. h) Place and seal the strip caps onto each strip tube using the GENE - UP® Lysis Tube Remover Tool. If less than eight caps are required, the caps can be cut apart, and only the used caps are placed onto the strip tubes.

10

Made with FlippingBook - Online catalogs