Table 1. Calibration and verification of equipment or systems
Equipment /System
Action
Frequency
Microscopes used for measuring
Calibrate stage micrometer
At installation or initial use
pH meters ion selective, and related conductivity equipment
Calibrate against reference buffer E or reference solution at level of use or bracketing range of use
Each day of use
Temperature controlled chambers used for storage (e.g., refrigerators and freezers)
Verify temperature
Frequency is dependent upon mechanism of monitoring F . At installation (or initial use) and after non-routine maintenance G Minimum twice daily on each day of use with at least 4 hours between verifications At installation (or initial use) and after non-routine maintenance G
Uniformity of temperature by mapping the chamber
Temperature controlled chambers used for testing (e.g., ovens, furnaces, incubators, water baths, and autoclaves)
Verify temperature
Uniformity of temperature A by mapping the chamber Calibrate temperature to the appropriate traceable standard
Temperature sensing devices/systems, reference (e.g., thermometers, thermocouples, data loggers, data tracers, thermistors, digital displays, continuous monitors, etc.) Temperature sensing devices/systems, working (e.g., thermometers, thermocouples, data loggers, data tracers, thermistors, digital displays, continuous monitors, etc.)
Every 2 years
Verify temperature against reference device
Annually
Verify working device annually against reference or against NIST time clock I
Timers and Internal Timing Devices H
Time
UV/Vis Spectrophotometer
Verify blank reading
Daily when in use
Verify wavelength
At installation and annually