Table 1. Calibration and verification of equipment or systems
Equipment /System
Action
Frequency
Volumetric delivery devices: mechanical pipets, mechanical burets, and liquid dispensers
Verify accuracy using mass of water at a known temperature or by spectrophotometric method
Upon receipt (manufacturer's Certificate of Accuracy may be accepted) Minimum every 6 months
Volumetric delivery devices: positive displacement syringes used for volumetric delivery
Verify accuracy
Upon receipt (manufacturer's Certificate of Accuracy may be accepted)
Volumetric non-class A glassware: pipets, burets, and volumetric flasks
Accuracy using mass of water at a known temperature or by spectrophotometric method Verify water activity of known solutions Method specific water quality attributes.
Upon receipt
Daily when in use E
Water activity meter
Minimally every month
Water, used in all analyses to meet method requirements
Water, used for microbiological analyses
Acceptable levels of Chlorine and Aerobic Plate count
Monthly