- Assay controls within acceptance criteria or specification? - System suitability parameters within acceptance criteria?
Laboratory Investigation Conclusion: a./ Assignable Cause: Initial results invalid, repeat test, retain the initial data, but do not use b./ No Assignable Cause: Inconclusive, retest the sample
4. Equipment qualification
Detector response (UV, DAD, Fluorescence, EI, RI, ELSD, MS)
Verify detector response/linearity with at least 3 concentrations of analytical standard or certified reference standard if available. Concentrations should include the low, high and the mid-range of the calibration curve. Pump flow accuracy and precision, column temperature accuracy and stability, wavelength accuracy, signal-noise and drift, injection precision, injection carry-over, gradient composition accuracy, gradient composition noise and drift, sample temperature accuracy
At least once with each batch
Chromatographic Systems (GC, IC, LC)
Annually, or after maintenance and repair