Title : Use of AFFINIMIP®SPE Bisphenols as clean up method for the determination of Bisphenol A from COLA drinks by Fluorescence detection Author(s): Kaynoush Naraghi, Sami Bayoudh, Michel Arotçaréna Submitted by: Michel Arotçaréna, AFFINISEP Primary Reviewer: Melissa Phillips, NIST

MOTION NOT to move this method to First Action Official Methods of Analysis Status (Ackerman/Seipelt) 9 in favor, 0 opposed, 2 abstentions. The motion passed. Reasoning/Summary: The ERP did not see data on any of the matrices besides cola drinks. There was no safety information. There was no calibration information. Energy drinks did not meet the SMPR. There was concern about the capacity of the column. ____________________________________________________________ Further Recommendations / Requirements: None .

BPA-03

Secondary Reviewer: Tom Seipelt, Abbott