OMA Consulting Protocol Review: OMACON-2019-2

2.4.4.1 Fill the measurement cuvette with 10 mL of Measurement Solution and put it in the cuvette holder making sure that all the components of the electrochemical cell are submerged. 2.4.4.2 Connect the device to a 100-240 VAC socket and press the “On” switch that is beside the screen for 2 seconds. 2.4.4.3 In the BIOTEST option: select the parameter to be analyzed. Press », enter the Biotest code (this code is on the Biotest ID label), press » and finally select the Measurement Range to be used and press Save.

2.4.4.4 The ANALYSIS option takes you to the main analysis menu: Activate, Calibrate, Standard and Sample.