RI-ERP-FINALACTION-Recommendations
OMAMAN-20 C/ In House Validation Report ERP Use Only - March 2015
Table 16. Results for testing a gliadin-contaminated stainless steel surface with an area of 10 x 10 cm for each swabbing experiment; 20 replicates per amount
µg gliadin/100 cm 2
result for each repeat
POD
- - - - - -
- - - - -
- - - - -
- - - - - - - -
-
-
0,00
0,00
- -
+
+ -
+ -
0,25
0,25
+ -
- -
-
-
-
+
+ + -
-
+ -
+ + -
0,50
0,25
- -
- -
- -
- -
-
- + + + + + + + + + + + + + + - - - - + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + +
1,00
0,80
2,00
1,00
4,00
1,00
The results of these experiments are given in table 16. This is the only “matrix” were a higher degree of variation occurs. Besides no false-positives, the concentration range with negative and positive results was between 0.25 µg and 1.0 µg gliadin per 100 cm 2 surface. Starting with 2 µg gliadin/100 cm 2 , all results were positive. The testing of surfaces was the only case where two invalid results occurred (one at 0.5 µg/100 cm 2 ; one at 1 µg/100 cm 2 ). These two were tested again in a blind-coded manner.
AOAC Research Institute Expert Review Panel Use Only
Figure 4. Probabilities of detection (POD) for surface stainless steel (swabbing procedure); 20 replicates per gliadin amount.
Page 25 of 27