Enzymatic HPLC Method
– Modifications to method that may improve performance/reproducibility between labs I i i f Δdi 0S f id k • ncreas ng retent on o ‐ away rom vo pea (several labs did not achieve resolution) – USP method uses anion exchange (SAX) column • Performing reaction in 2‐ mL or 5‐ mL conical bottomed reaction vials to eliminate transfer step • For raw materials, dry sample in oven prior to weighing, and place desiccant in balance chamber
Concerns Surrounding Enzymatic HPLC Method – Lack of availability and purity issues associated with individual disaccharide reference standards • Adopting USP’s method of using a USP Chondroitin – Limited supply of enzyme • IBEX was only supplier for the past couple of years but recently Sigma started offering Chondroitinase AC again so availability is no longer an issue – Expensive to run – Technically Challenging Sulfate RM (or other well characterized/qualified Chondroitin RM) verses the individual disaccharides would address this issue