Analytical Range: 0.2 to 4.2 mg/L in measured solutions but could not be determined per AOAC. LOQ: 0.2 mg/L Accuracy / Recovery: Recovery was 97%. Could not be ascertained in the absence of cis‐isomer standards. Precision: RDS was 0.23%. Reproducibility: RSD <0.92% Conditions not provided to allow suitability to be replicated. —All instrument configuration; instrument details, including serial numbers and acquisition information; and column details, including identification, were properly recorded during laboratory operations. Equipment calibration, including signal description and baseline noise determination, was performed before tests were executed. This is the basis of a functional method. Several things could improve and expand the method. It may be more applicable to measuring free lutein and zeaxanthin than the esters. I would recommend that some things be adjusted before recommending first action status. Sample prep to address additional products. Assessment of the composite‐specific absorbance. Inclusion of standards and beta‐cryptoxanthin.
Performance Characteristics
System Suitability
Do you recommend this method be adopted as a First Action and published in the Official Methods of Analysis of AOAC INTERNATIONAL? Please specify rationale. Recommendation for the Method