SPDS Meeting Book
SMPR Key Points
1. Method Performance Requirements :
Table 1: Analytical Range & LOQ Based on Matrix Parameter Finished Products
Raw Materials
Analytical range ppm*
0.5 – 338
1,250 ‐ 12,500
Limit of Quantitation ppm*
0.4
1,000
Table 2: Method Performance Requirements as a Function of Range
Ranges (µg/g)*
Parameter
< 10 ‐ 15
>15 ‐ 50
>50 – 500
>500 – 4,000
>4000 – 12,500
Recovery (%) % R t bilit epea a y (RSD r ) % Reproducibility (RSD R )
80 ‐ 110
90 ‐ 107
95 – 105
95 – 105
97 – 103
8
7
5
4
3
12
10
8
6
4
* Measured as individual forms of Vitamin D and pre‐Vitamin D
Comments Submitted (if any)
• One Comment Submitted – “The two certified reference materials I
mentioned were vitamin D2 and vitamin D3. We have other vitamin D standards in the attached but they are the metabolites.” • Cerilliant Vitamin D2V‐024 Vitamin D3V‐025 – 1 mg/ml CRM in Ethanol St k h ld D dd t SMPR – a e o ers: o we a o as a requirement to use?