SPDS Set 1 ERP Book

ANTH-002

LOQ:

Not covered.

Accuracy/Recovery

Not covered.

Precision (RSDr)

Not covered.

Reproducibility (RSDR)

Not covered.

System Suitability

Not covered.

Safety Review

Is the method safe to use as written? Are there any safety concerns with any component, instrument, reagent or instruction of the method? If yes, please clearly state the concerns. Does the method include all necessary safety precautions and/or warning statements? Yes or No. If No, please suggest potential precautions or warnings that can be added to the method for users or recommend safer option(s) if warranted.

Yes. There are no safety precautions outlined in this method. At a minimum, the use and handling of perchloric acid as well as 12N HCl needs to be covered.

No

Recommendations

No. The method does not address 7 of the 8 Vaccinium as outlined in the SMPR Annex 1. It is not clear that this method would be able to differentiate/identify the various species or identify adulterant presence/absence as required by the SMPR.

Final Recommendation:

Made with