FOS-04 FOR ERP USE ONLY DO NOT DISTRIBUTE
Figure 2-3 Graph of Standard Calibration Errors
2. Accuracy – Accuracy was determined by testing the spiked samples described in Table 2-2. Mean recoveries ranged from 96- 102%. The results are summarized in Table 2-5.
Table 2-5 Accuracy/Spike Results
0.03% FOS spike (9% sucrose)
0.3% FOS spike
0.3% P95 spike
0.3% Synergy 1 spike
3% FOS spike
3% P95 spike
3% Synergy 1 spike
min% recovery avg % recovery max % recovery
90.8%
91.7%
93.9%
94.1%
92.6%
92.5%
95.3%
102%
100%
99.0%
98.4%
96.2%
96.0%
97.2%
115%
107%
113%
104%
101%
99.0%
101%
%RSD
7.12%
5.43%
6.09%
2.85%
2.85%
2.35%
2.04%
0.03%, 0.3%, and 3% are nominal spike values, N=9 (three replicates in each of three independent runs) for all levels and fructans.
3. Precision – Precision was assessed by both the analysis of 20 independent duplicate sample preparations (using 6-7 analysts and 4-5 instruments in 2-3 labs) as well as the overall assessment of additional analyses of the AN internal control (using 7 analysts and 9 instruments in 3 labs). a. Precision was additionally assessed for 6 SPIFAN matrices by four days of duplicate testing (Table 2-9). Data is still being collected at this time.