SPSFAM Meeting eBook March 2017
3/9/2017
SMPR Key Points
Concentrates
Dried Plant Material
Requirement (Additional, Desirable Cannabinoids (Table 1b), including CBN)
Requirement (Pertains to Cannabinoids in Table 1b)
Parameter
Requirement THC, THCA, CBDA, CBD, Individually Reported
Parameter
Limit of Quantitation (LOQ) (%)
≤ 0.1
Limit of Quantitation (LOQ) (% w/w)
≤ 0.3
≤ 0.3
Analytical Range (%)
0.1 – ca. 50*
Analytical Range (% w/w)
≤ 0.3 – ca. 100
≤ 0.3 – ca. 50
*Lower concentrations may be acceptable as applicable for cannabinoids listed in Table 1B.
Ranges (%) (Pertains to Cannabinoids in Table 1b)
Ranges (% w/w)
Parameters
Parameters
≤ 0.3 – 1
> 1 ‐ 10 > 10 – ca. 100
0.1 – 1
> 1 ‐ 25 > 25 – ca. 50
95 – 105
97 ‐ 103 98 ‐ 102
Recovery (% w/w)
Recovery (%)
95 – 105
97 ‐ 103
98 ‐ 102
≤ 5
≤ 4
≤ 2
% RSD r
% RSD r
≤ 5
≤ 4
≤ 2
≤ 7
≤ 5
≤ 3
% RSD R
% RSD R
≤ 7
≤ 5
≤ 3
Comments Submitted
5 Comments were received for SMPR –dried materials. 1. The proposed procedures should add (a) Instrument Detection Limit (IDL), and (b) Method Detection Limit (MDL) Action: WG decided to use LOD which is the equivalent of IDL, which uses blank matrix samples; LOD uses matrix samples. MDL was purposefully removed by the WG.
7