SPSFAM Meeting eBook March 2017
93 94
Table 2: Method performance requirements (part 1).
Requirement THC, THCA, CBDA, CBD, Individually Reported
Requirement (Additional, Desirable Cannabinoids (Table 1b), including CBN)
Parameter
Limit of Quantitation (LOQ) (% w/w)
≤ 0.3
≤ 0.3
Analytical Range (% w/w)
≤ 0.3 – ca. 100
≤ 0.3 – ca. 50
95 96 97 98 99
*Reported as individual cannabinoids
100 101 102 103 104 105 106
Table 3: Method performance requirements (part 2).
Ranges (% w/w)
Parameters
≤ 0.3 – 1
> 1 - 10
> 10 – ca. 100
Recovery (% w/w)
95 – 105
97 - 103
98 - 102
% RSD r
≤ 5 ≤ 7
≤ 4 ≤ 5
≤ 2 ≤ 3
% RSD R
Made with FlippingBook