pC metamer: Report
AOAC International Interlaboratory Study Workbook
Version:
4.1
Blind (Unmatched) Replicates Study Reported Values
Seq.
Item
Symbol
Value 95% LCL 95% UCL
Name of study: Date of study: Sample/Matrix ID: Analyte:
JAOAC (2012) 95(2):411-418
1-Mar-2013
C02
OBPCP
Factor for Units of measurement: Metamer to analyze (C or pC): True value (blank if not known):
0.01
pC
1 Total number of laboratories 2 Total number of replicates
p
13 26
Sum(n(L)) XBARBAR
3 Overall mean of all data (grand mean) 4 Repeatability standard deviation 5 Interlaboratory standard deivation 6 Bias estimate (Measured - True) DERIVED VALUES: 7 Reproducibility standard deviation 8 Root-mean-square-total-error 9 Repeatability relative standard deviation 10 Reproducibility relative standard deviation
3.0226 3.0087 3.0364 0.00563 0.00409 0.00908 0.02370 0.01699 0.03912
s(r) s(L)
b
s(R) rmse
0.02436
RSD(r) RSD(R)
11 HORRAT value
(RSD(r) and RSD(R) are not appropriate for pC data, as they correspond directly to s(r) and s(R) here.)
17
copyright 2018 by Least Cost Formulations, Ltd., all rights reserved.
Compare results for C and pC metamers
Metamer Quantity Value
RSD(r)
RSD(R)
C
1.34% 5.52%
pC pC
s(r)
0.00563 1.30%
s(R)
0.02370
5.46%
• RSD(r) for C metamer = 2.303 x s(r) for pC metamer. • RSD(R) for C metamer = 2.303 x s(R) for pC metamer. • No need for RSD(r) and RSD(R) if fit using pC metamer. • RSD(r) and RSD(R) from pC metamer values are unbiased and efficient estimators for these quantities, unlike those derived from C metamer values. • 2.303 = ln(10), an artifact of using log 10 instead of ln