Total Collaborative Study Protocol_Solus One Salmonella v1 1

the kit on a sanitized bench top for 1 h prior to use. 

( b ) Determine the number of wells required for the test. Take the required number of strips from  the pouch and fit them to the frame provided. Note: Unused strips should be returned to the pouch and 

stored at 2–8°C. 

Analysis 

( a )  Prepare equipment 1) Turn on power to the Dynex DS2 instrument, followed by the power to the computer. 

Open the Matrix software on the computer and enter login information. The Dynex DS2 

instrument will perform a self‐diagnostic check. 

( b )  Create a Run File 

1) In the DS‐ Matrix software, select “File”, “Worklist Editor”. Then select “New Sample 

Batch”. 

2) Enter your sample batch ID as required. 

3) Enter the total number of samples in the batch. 

4) Sample ID may then be typed into the empty positions in the rack or assigned 

automatically using an onboard barcode scanner. 

5) If using an onboard barcode scanner slide the sample tube racks onto the machine and 

click next. 

6) Click on the Solus One  Salmonella assay. Then click the “Add Assay” button.  

7) Select “Done” when assay selection is completed. 

8) Select the “Accept” button to start the run.