Total Collaborative Study Protocol_Solus One Salmonella v1 1
MALDI Biotyper biochemical identification method, AOAC OMA 2017.09 (9).
Comparison of the candidate Solus One Salmonella and reference methods for all matrices indicated
un‐inoculated test portions were all negative, high inoculation level test portions were all positive and
fractional levels were achieved at the low inoculation level test portions using an automated sample
preparation method.
As per criteria outlined in Appendix J of the Official Methods of Analysis Manual, fractional positive
results were obtained for each matrix (3). The POD analysis between the candidate Solus One
Salmonella method and the FDA/BAM Ch. 5 reference method for all matrices indicated that there was
no significant difference at the 5% level between the numbers of positive results by the two methods
using an automated sample preparation method.
The POD analysis between Solus One Salmonella presumptive and confirmed results for all matrices
indicated that there was no significant difference at the 5% level using an automated sample
preparation method. A summary of POD analyses is presented in Tables 3 and 4.