2016 OMB Summer Meeting
AOAC OMB Meeting Book
393
AOAC 2012.22 (FINAL VERSION REVIEW) - JUNE 2016 FOR ERP USE ONLY DO NOT DISTRIBUTE
77
1. Injection volume.- 5µL.
78
2. Autosampler temperature.- 10 °C.
79
3. Column temperature.- 25 °C.
80
4. Flow rate.- 0.35mL/min
81
5. Run time.- 4.0 minutes
82
6. Mobile phase.- Sodium acetate (50 mmol/L, pH 5.4), decylamine (1.6g),
83
acetonitrile (1 %), TCEP (40 mg/L).
NOTE: At the end of each analytical series, rinse column with acetonitrile-water (1+1, v/v) 84
for 10 minutes at 0.4 ml/min .
85
(b). System suitability test .- Equilibrate the chromatographic system for ≥0.5 h.
86
87
Inject a working standard solution of ascorbic acid at least 6 times and check
88
peak retention times and response (peak height or area). Inject working
89
standard solutions on a regular basis within a series of analyses. Ensure orotic
90
acid and isoascorbic acid are fully resolved from ascorbic acid by injecting
91
separated standard solutions of both compounds (prepared as stated for
92
ascorbic acid). If not resolved, the pH of the mobile phase may be decreased to
93
5.0, or the amount of acetonitrile increased.
(c). Calibration .- Make single injections of working standard solutions at least at
94
95
the beginning and the end of each analytical series. Establish the calibration
96
curve (seven points) by plotting peak response (height or area) versus ascorbic
97
acid concentration. Perform linear regression. Calculate the slope (S) and
98
intercept (I) of the calibration curve.
(d). Analysis .- Make single injections of sample solutions.
99