AOAC CASP Meeting - MYM 2020
DRAFT AOAC SMPR 2019.XXX; Version 6; March 11, 2020. 1 2 Method Name: 3
Determination of Heavy Metals in a Variety of Cannabis
4 5 6 7 8 9
and Cannabis Derived Products
Approved by:
Final version date : Effective date:
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55
Intended Use :
Surveillance methods for routine monitoring
1. Applicability :
Determination of total cadmium (CAS No. 7440-43-9), total arsenic (CAS No. 7440- 38-2), total lead (CAS No. 7439-92-1), and total mercury (CAS No. 7439-97-6).
Additional elements in Table 1 may be included.
2. Analytical Technique :
Inductively Coupled Plasma based instrumentation or alternative methodology that
meets the performance requirements
3. Definitions :
Limit of Quantitation (LOQ)
The minimum concentration or mass of analyte in a given matrix that can be reported
as a quantitative result.
Repeatability
Variation arising when all efforts are made to keep conditions constant by using the same instrument and operator, and repeating during a short time period. Expressed as the repeatability standard deviation (SD r ); or % repeatability relative standard
deviation (%RSD r ).
Reproducibility
The standard deviation or relative standard deviation calculated from among- laboratory data. Expressed as the reproducibility relative standard deviation (SD R ); or
% reproducibility relative standard deviation (% RSD R ).
Recovery
The fraction or percentage of spiked analyte that is recovered when the test sample
is analyzed using the entire method.