AOAC CASP Meeting
Method Performance Requirements Table 2: Method performance requirements (part 1) for cannabinoids
Parameter *
Requirement
Limit of Quantitation (LOQ) (%) Analytical Range (CBD & CBDA) (%)
≤ 0.05
0. 05 – 35
Analytical Range (others) (%)
0.05 – 5
*All calculated on dry weight basis
Table 3: Method performance requirements (part 2) for cannabinoids Method Performance Requirements
Ranges (%)
Parameters*
0.05 – 0.5
> 0.5 ‐ 5
5 ‐ 35
Recovery (%)
85 – 118
90 ‐ 111
95‐105
% RSD r % RSD R
≤ 5
≤ 3
≤ 2
≤ 8
≤ 6
≤ 4
*All calculated on dry weight basis; observed values to be compared to indicated limits for acceptability.