The General MS parameters for API 5500 (Qtrap): Curtain gas(nitrogen):
25 psi;
Collision Assisted Dissociation
Medium;
IonSpray voltage:
5500 V;
Temperature:
700 °C;
Ion source gas1(zero air):
60 psi;
Ion source gas2(zero air):
50 psi;
Platform: Agilent 1290 Infinity Binary LC with Sciex 6500 (Q trap) LC- MS/MS system or 6500 Plus LC-MS/MS system: HPLC Conditions : Agilent 1290 Infinity Binary LC System: LC Column: Waters ACQUITY UPLC HSS T3 Column (100Å, 1.8 µm, 2.1 mm X 50 mm). Mobile Phase: A: 1% Glacial acetic acid and B: MS grade methanol as mobile phase components in different ratios at different time of the analysis. The Pumping gradient of two solvents at different time is as the following.
Flow (mL/min)
Time (min)
%B
0
0
0.5
0.5
0
0.5
5.00
40
0.5
5.05
100
0.5
6.50
100
0.5
6.55
0
0.5
9.00
0
0.5
A = 1% Glacial Acetic Acid in Water, B = 100% Methanol
Column Temperature: 40 ° C
Collaborative Study Report: Method AOAC 2011.06 (Folate by LC-MS/MS) 40