AOAC SPIFAN Nutrients ERP-Final Review (Oct. 2018) Pubs
2011.06 (September 2017)-Fol-22 MLT Report
FOR ERP USE ONLY DO NOT DISTRIBUTE
Autosampler Temperature: 5 ° C
Injection Solvent: Std., samples and blanks injection solvent is 90% MeOH + 10% Formic acid + 1% ascorbic acid neutralized to pH 7 with ammonium hydroxide (28-30%), 50 uL/mL. Injection Volume = 2 uL MS Conditions for the Sciex 6500 LC-MS/MS (Qtrap): MS conditions Sciex 6500 for detection of different transitions of the analytes in the scope Comp ound Q1 Q3 DP EP CXP CE
442.2
295.2
75
6
6
20
FA
442.2
176.2
75
6
6
45
460.2
313.2
75
6
6
28
5-CH3-THF
460.2
194.2
75
6
6
46
474.2
327.2
75
6
6
25
5-CHO-THF
474.2
299.2
75
6
6
40
13 C 5
-FA
447.2
295.2
75
6
6
20
13 C 5
-5-CH3-THF 465.2
313.2
75
6
6
28
13 C 5
-5-CHO- THF
75
6
6
479.2
327.2
25
DP: Declustering potential; EP Entrance potential; CEP collision cell entrance potential; CE Collision cell; CXP collision cell exit potential. The General MS parameters for the Sciex 6500 LC-MS/MS (Qtrap):