Table A2: Expected Precision as a Function of Concentration-Repeatability
Analyte (%)
Analyte Ratio
Unit (%, mg/kg)
%RSD
% Mean Recovery
100
1.0
100 %
1.3 1.9 2.7 3.7 5.3 7.3
98-102 98-102 97-103 95-105 90-107 80-110 80-110 80-110 60-115 40-120
10
10 -1 10 -2 10 -3 10 -4 10 -5 10 -6 10 -7 10 -8 10 -9
10 %
1
1 %
0.01
0.1 %
0.001
100 mg/kg (ppm) 10 mg/kg (ppm) 1 mg/kg (ppm) 100 g/kg (ppb) 10 g/kg (ppb) 1 g/kg (1ppb)
0.0001 0.00001 0.000001 0.0000001 0.00000001
11 15 21 30
9
Definitions Multi-laboratory Validation (MLV) — Demonstration between laboratories using adulterated samples created by a third-party group and supplied blindly to the participating laboratories according to guidelines described by the AOACI [Reference Appendix Document] to be considered for classification as AOAC Final Action Method " Protocol for the design, conduct and interpretation of method performance studies ". Pure and Applied Chemistry, Horwitz, W. 1995. 67:331-343; " Guidelines for the Assessment of the Competence of Testing Laboratories Involved in the Import and Export Control of Food“- CAC/GL 27-1997 ; " Harmonized IUPAC Guidelines for the use of Recovery Information in Analytical Measurement" - CAC/GL 37-2001 ; see also " Harmonised Guidelines for the Use of Recovery Information in Analytical Measurement" Thompson, M., Ellison, S., Fajgelj, A., Willetts, P., & Wood, R. (1999) Pure Applied Chemistry, 71: 337-348.