Specific performance criteria
Table 2: Mass spectrometric performance and peak identification criteria
(U)HPLC-LRMS
Unit mass resolution
Typical systems (examples)
MS/MS triple quadrupole, ion trap, Q-trap, Q-TOF, Q-Orbitrap
Acquisition
Selected or multiple reaction monitoring (SRM, MRM)
Minimum number of ions
2 product ions
Ion ratio from sample extracts should be within ± 30 % (relative) of average of calibration standards from same sequence.
Ion ratio
Signal to noise (S/N) ratio
≥ 3
The ratio of the chromatographic RT of the analyte to that of the IS (i.e. relative RT of the analyte) shall correspond to that of the calibration standard with a maximum deviation of 1 %. Analyte peaks from both product ions in the extracted ion chromatograms must fully overlap.
Retention time (RT)
Other
AOAC Midyear Meeting - Overview of EURL PFAS Guidance Document
16.03.2023