Review Team (February 2016)
Amino-03 (February 2016) FOR ERP USE ONLY DO NOT DISTRIBUTE
Amino Acids in Infant Formula and Adult/Pediatric Formulas Ultra High Performance Liquid Chromatography
Column
ACCQ-TAG ULTRA C18 1.7μM, 2.1 x 100mm
Mobile phase A Mobile phase B
Mobile Phase A1 Mobile Phase B1 0.6 mL/minute
Flow rate
Injection Volume
1.0 μL 50 °C 260 nm
Column Temperature Detector Wavelength
Run Time
14 minutes Gradient Method
Time (minutes)
A %
B%
0.00 0.63 2.33 2.92 3.50 4.12 4.70 5.64 5.99 6.16 6.39 6.70 9.03
99.90 99.90 99.50 98.70 98.00 97.00 96.00 93.90 93.10 92.60 92.00 90.90 78.80 40.40 10.00 10.00 99.90 99.90
0.10 0.10 0.50 1.30 2.00 3.00 4.00 6.10 6.90 7.40 8.00 9.10
21.20 59.60 90.00 90.00
10.00 10.01 11.01 11.02 14.00
0.10
0.10 3) Chromatographic conditions for Alkaline Hydrolysis method: Table 8: Column
Agilent Zorbax RRHD, 1.8μm, 2.1 X 50 mm
Mobile phase
Mobile Phase A2 0.2 mL/minute
Flow rate
Injection Volume
2.0 μL
Column Temperature Detector Wavelength
25°C
Ex. 280nm, Em. 356nm
Run Time
10minutes
Pump Mode
Isocratic
7