Table 7. Overview performances Analyte Mn Cu

Fe

Zn

Mg

P

Na

Ca

K

Lower analytical range a Upper analytical range b Limit of Quantitation (LOQ) Repeatibility ≤ (RSD) c

0,003

0,010

0,2

0,01

0,09

0,04

0,5

0,4

6

5

5

125

25

375

1250

625

1875

1875

0,003

0,010

0,2

0,01

0,09

0,04

0,5

0,4

6

1,2%

0,5%

1,3%

1,8%

1,5%

1,0%

0,7%

1,7%

0,8%

Recovery d

97,4%

100,3%

96,9% 96,0% 97,6%

103,7%

95,5%

98,9%

98,5%

Reproducibility ≤ (RSD) e

1,4%

1,7%

0,9%

1,0%

1,2%

1,0%

1,2%

1,2%

1,4%

a Lower analytical range is equivalent to the LOQ. b Upper analytical range is equivalent to the highest standard (5) of the calibration curve used on daily basis calculated by multiplying this standard by a dilution factor of 25. c Repeatibility of NIST SRM 1849a calculated by using the residues of the duplicates on 9 different days. d Overall average for 3 independent days using NIST SRM 1849a and the laboratory control sample CM01/07. e Reproducibility of NIST SRM 1849a calculated by using the average results of the duplicates on 9 different days. *mg/100g reconstituted final product.