BVit-02 - MEF FORM FOR ERP USE ONLY DO NOT DISTRIBUTE
Table 2
Accuracy - Analytical bias compared to reference material Certified Reference Material
Measured Results
T stat Bias
Difference from certified value
CRM product
Analyte
Certified value Uncertainty Certified range Concentration
Confidence interval
p -value
Nominal
990 987
90.1 92.8 38.8 76.0
1041 1004 1074 1067 1079 1117
6.3
13.3
SRM 1849a
Niacin (ug/100g)
1080
100
980 - 1180
-50.3
3.792
0.003
1000.5 - 1058.9
1.4
-37.3 100.9 103.1
979 977
1030 1012
49.8 68.5
Mean
1029.7
50.3
SD
46.0
46.0
d
=
T stat
/ sd
n
where: = mean of the differences between measured results and reference value sd = standard deviation of the differences between results and reference value n = number of replicate analyses of SRM d
H 0 : = 0, The null hypothesis is that there is no difference between the measured results and the certified value: d
p -value: is the probability (at the 95% levels of confidence) of obtaining the test statistic assuming that the null hypothesis is true. p -value < 0.05 indicate a difference of statistical signifance between measured results and SRM.
95% confidence interval: The true population of the measured value expected to fall within the confidence interval estimate 95% of the time.
sd
sd
t crit
t crit
lower limit = mean -
upper limit = mean +
n
n
t
where: = critical values of the t-distribution
crit
sd = standard deviation of the differences between results an reference value n = number of replicate analyses of SRM
SPIFAN ERP Checklist v1.0