Table 2
Accuracy - Analytical bias compared to reference material Certified Reference Material
Measured Results
T stat Bias
Difference from certified value
CRM product
Analyte
Certified value Uncertainty Certified range Concentration
Confidence interval
p -value
Nominal
131.7 134.1 129.9 130.5 129.7 126.2 143.3 144.3 130.4 129.3 129.5 128.6
2.9 0.5 4.7 4.1 4.9 8.4
SRM 1849a
Pyridoxine (ug/100g)
134.6
9.3
125.3 - 143.9
-2.3
1.406
0.187
128.7 - 135.9
-8.7 -9.7
4.2 5.3 5.1 6.0
Mean
132.3
2.3
SD
5.7
5.7
d
=
T stat
/ sd
n
where: = mean of the differences between measured results and reference value sd = standard deviation of the differences between results and reference value n = number of replicate analyses of SRM d
H 0 : = 0, The null hypothesis is that there is no difference between the measured results and the certified value: d
p -value: is the probability (at the 95% levels of confidence) of obtaining the test statistic assuming that the null hypothesis is true. p -value < 0.05 indicate a difference of statistical signifance between measured results and SRM.
95% confidence interval: The true population of the measured value expected to fall within the confidence interval estimate 95% of the time.
sd
sd
t crit
t crit
lower limit = mean -
upper limit = mean +
n
n
t
where: = critical values of the t-distribution
crit
sd = standard deviation of the differences between results an reference value n = number of replicate analyses of SRM